SUPER GL compact

The SUPER GL compact has the advantages of SUPER GL easy+ and SUPER GL ambulance in one device.

With this analyser it is possible to measure glucose, lactate and hemoglobin simultaneous from one sample and the device also shows the glucose values plasma referenced. At it’s sample rotor are one position for calibration, one position for sample measuring and two positions for control material. Additional the device has a sample rack with six sample positions for patients samples - so you also could measure small series.

The software of the device also calculates the statistical evaluation of the control measuring according the actual valid German guideline of the German Medical Association.This statistical evaluation is easy to program and easy to use by the innovative touchscreen. The statistical evaluation of the controls will be shown at the display, but also could be printed or send to EDP.

A further news at this product generation is the entry of the barcode by a PS2-keyboard or a barcode reader - and so its much easier to connect the measured value to the patients data.

We realised the IFCC-recommendation for displaying the glucose values plasma referenced at this device.

Please note that the order form is not available via the mobile version. Alternatively, you can use the order form as a PDF.

Performance Data

Measuring Time

Single Measurement Glucose ca. 30 seconds
Single Measurement Glucose / Lactate / Hemoglobin ca. 45 seconds
Series Measurement Glucose ca. 30 seconds - 120 samples/hour
Series Measurement Glucose / Lactate / Hemoglobin ca. 45 seconds - 80 samples/hour

Measuring Range

Glucose 0.5 mmol/l to 50 mmol/l or 9.01 mg/dl to 900.8 mg/dl
Lactate 0.5 mmol/l to 30 mmol/l or 4.5 mg/dl to 270 mg/dl
Hemoglobin 1.2 mmol/l to 20 mmol/l or 1.9 g/dl to 32 g/dl

Sample Volume

10 µl sample diluted with 500 µl hemolysate solution

Precision (18 samples) - Variation Coefficient Indication

Glucose (12 mmol/l or 216 mg/dl) < 1.5 %
Lactate (10 mmol/l or 90 mg/dl) < 1.5 %
Hemoglobin (7.6 mmol/l or 12 g/dl) < 1.5 %

Sensors

Operating Time Glucose Sensor max. 6 months / max. 10.000 samples
Operating Time Glucose and Lactate Sensors max. 3 months / max. 10.000 samples

Technical Specifications

Working Temperatur + 15 °C bis + 35 °C
Supply Voltage 12 V DC
Power Consumption max. 45 W
Classification according MPG In-vitro-Diagnosticum (guidline 98/79/EC)

Interfaces

Printer serial, RS 232
EDP serial, RS 232
Barcode Reader external, PS/2

Dimensions / Weight

Width 220 mm
Hight 170 mm
Depth 185 mm
Weight 1.7 kg
General data
Sensors
QuantityItem numberItem descriptionDevice(s)
Washing solution
QuantityItem numberItem descriptionDevice(s)
Reagent cartidges
QuantityItem numberItem descriptionDevice(s)
Hemolysate-System-Solution
QuantityItem numberItem descriptionDevice(s)
Control material
QuantityItem numberItem descriptionDevice(s)
Glucocapil reaction cups
QuantityItem numberItem descriptionDevice(s)
Capillaries
QuantityItem numberItem descriptionDevice(s)
Glucocapil calibration solution
QuantityItem numberItem descriptionDevice(s)
Miscellaneous
QuantityItem numberItem descriptionDevice(s)
Sets
QuantityItem numberItem descriptionDevice(s)
Print order form

Documents

Declarations of conformity